分析与监测

水中铅的原子荧光光谱测定法

展开
  • 广东省肇庆市自来水公司, 广东肇庆 526020
谢勇坚(1967-),1989年毕业于广东省韶关卫生学校,工程师.电话:0758-2852014.

网络出版日期: 2010-10-01

Determination of lead in water by atomic fluorescence spectrometry

Expand
  • Zhaoqing City Water Supply Co., Zhaoqing 526020, China

Online published: 2010-10-01

摘要

采用原子荧光光谱法测定水中铅。实验结果表明,线性关系良好(在标准曲线质量浓度为0.00~10.00μg/L内,平均相关系数γ为0.9990),精密度较好(相对标准偏差为0.96%~3.52%),准确度、灵敏度较高(加标回收率为90.3%~108.4%,最低检出限为0.09μg/L),是测定水中铅的较好方法。

本文引用格式

谢勇坚 . 水中铅的原子荧光光谱测定法[J]. 工业水处理, 2004 , 24(6) : 50 -51 . DOI: 10.11894/1005-829x.2004.24(6).50

Abstract

Atomic fluorescence spectrometric determination of lead in water is carried out. The results obtained show that linearity is good(the related average coefficient is 0.999 0 in standard curve 0.00 ~ 10.00 μg/L). The precision is better(the relative standard deviation is 0.96% ~ 3.52%). The accuracy and sensitivity are higher(the recovery rate is 90.3% ~ 108.4%, detection limit is 0.09 μg/L). It is a pretty good method for the determination of lead in water.

参考文献

[1] 中华人民共和国卫生部卫生法制与肤督司. 生活饮用水卫生规范[M]. 北京: 中华人民共和国卫生部, 2001. 173-177
[2] 魏复盛, 等. 水和废水监测分析方法指南(下册)[M]. 北京: 中国环境科学出版社, 1997. 24-25