煤化工废水中无机阴离子的检测方法研究
|
|
高健,尚云涛,张洪杰
|
Study on detection of inorganic anions in coal chemical wastewater
|
|
Jian Gao,Yuntao Shang,Hongjie Zhang
|
|
表5 加标回收率实验
|
| |
|
| 阴离子 | 本底质量浓度/(mg·L-1) | 加标质量浓度/(mg·L-1) | 检测质量浓度/(mg·L-1) | 加标回收率/% | | F- | 3.75 | 2.00 | 5.85 | 102.67 | | 3.88 | 2.00 | 5.73 | 96.13 | | 3.61 | 2.00 | 5.71 | 102.77 | | Cl- | 334.59 | 100.00 | 437.68 | 100.92 | | 337.82 | 100.00 | 442.74 | 101.46 | | 336.21 | 100.00 | 435.11 | 99.67 | | NO2- | 63.18 | 10.00 | 74.29 | 101.76 | | 60.27 | 10.00 | 69.24 | 98.29 | | 64.81 | 10.00 | 72.88 | 97.02 | | Br- | 22.86 | 10.00 | 31.84 | 95.54 | | 24.72 | 10.00 | 35.73 | 104.09 | | 26.83 | 10.00 | 35.96 | 96.76 | | NO3- | 110.43 | 100.00 | 212.62 | 101.98 | | 113.65 | 100.00 | 212.35 | 98.86 | | 115.38 | 100.00 | 214.64 | 99.36 | | PO43- | 394.87 | 100.00 | 495.21 | 100.09 | | 396.14 | 100.00 | 495.28 | 99.78 | | 397.22 | 100.00 | 499.41 | 100.55 | | SO42- | 366.73 | 100.00 | 465.53 | 99.67 | | 368.87 | 100.00 | 472.19 | 100.90 | | 365.26 | 100.00 | 466.33 | 100.29 |
|
|
|