氯气校正法测定高氯水COD方法的标准化研究
王文雷, 宋永超, 刘金芝, 马振国, 王振华

Study on standardization of determination of COD in high chlorine water by chlorine emendation method
Wenlei WANG, Yongchao SONG, Jinzhi LIU, Zhenguo MA, Zhenhua WANG
表11 实际样品的加标测试结果
Table 11 Results of spiking test of actual samples
项目测定质量浓度/(mg·L-1加标质量浓度/(mg·L-1加标后测定质量浓度/(mg·L-1加标回收率/%平均加标回收率/%
地表水入海控制断面水样,Cl-为11 000 mg/L25.320.043.685.092.4
27.545.393.5
25.144.589.5
28.446.398.5
27.245.695.0
26.145.293.0
RSD/%4.92.15.1
项目测定质量浓度/(mg·L-1加标质量浓度/(mg·L-1加标后测定质量浓度/(mg·L-1加标回收率/%平均加标回收率/%

石化废水

水样,Cl-为8 200 mg/L

11610020194.094.3
11019891.0
10320396.0
9819790.0
11320699.0
10220396.0
RSD/%6.51.73.6