Study of the factors affecting the determination of totaliron content in circulating water by ophenanthroline

Expand
  • Research Institute of Beijing Yanshan Petrochemical Corp., Beijing 102500, China

Received date: 2001-08-19

  Online published: 2010-10-01

Abstract

The facto rs which affect the determination of to tal iron content in circulating water by o - phenan-throline are studied.the f actors are the pH, the vessels and the order of adding oxammonium hy drochloride, etc.the results show thathe pH must be controlled between 2~7, all vessels must be soaked in hydrochlo ricacid for 40 minutes.Add oxammonium hy drochloride before the ci rculating water is boiled.In this case.

Cite this article

CHENG Min, LI He-sheng . Study of the factors affecting the determination of totaliron content in circulating water by ophenanthroline[J]. Industrial Water Treatment, 2002 , 22(1) : 48 -51 . DOI: 1011894/1005-829x.2002.22(1).48

References

[1] 水和废水监测分析方法编委会.水和废水监测分析方法[M].北京:中国环境科学出版社,1990:180—182.
[2] 王镇浦,等.水质词汇、水质取样和分析方法[M].北京:水利电力出版社,1989:194.
[3] 张振中,等.化工环境监测实用手册[M].北京:中国科学技术出版社,1991:155—156.
[4] 谢贤群,王立军.水环境要素观测与分析[M].北京:中国标准出版社,1998:67.

Outlines

/