摘要:
采用普析通用公司生产的TW-6000自动水质监测仪测定COD,与标准法 (GB11914-1989)进行对照实验,再应用数理统计方法对结果加以分析,证明仪器法和标准法的数据无显著性差异 (显著性水平为0.05,样本数为15),且两者具有相似的精密度和准确度
关键词:
COD,
自动水质监测仪,
数理统计分析,
t检验
Abstract:
The water quality automonitor, TW- 6000, produced by Beijing Purkinje General Inst rument Co.,Ltd. is adopted to examine COD, to do check test with standard method ( GB11914 1989) and to analyze the results by mathemat ical stat ist ics. It is proved that there is no not iceable dif ference between the inst rumental method and the standard method(is 0.05 and the number of samples is 15) and the two methods have similar precision and accuracy.
Key words:
water quality automonitor,
mathemat ical analysis,
ttest
中图分类号:
郭强, 王文忠, 张涛, 王军. TW-6000 COD自动监测仪与标准法的对照实验[J]. 工业水处理, 2002, 22(7): 34-35.
GUO Qiang, WANG Wenzhong, ZHANG Tao, WANG Jun. Check test of TW- 6000 COD automonitor with standard method[J]. INDUSTRIAL WATER TREATMENT, 2002, 22(7): 34-35.