工业水处理 ›› 2009, Vol. 29 ›› Issue (7): 66-68. doi: 10.11894/1005-829x.2009.29.(7).66

• 分析与检测 • 上一篇    下一篇

连续流动分析仪测定水中阴离子洗涤剂的方法

汪言满   

  1. 中山市环境监测站, 广东中山 528403
  • 收稿日期:2009-03-20 修回日期:2009-03-20 出版日期:2009-07-20 发布日期:2010-10-01
  • 作者简介:汪言满(1972- ),2002年毕业于广东工业大学,硕士,工程师.电话:0760-88873210,E-mail:gdutwym@21cn.com

The determination of anionic detergent in water with continuous flowing analyser

Wang Yanman   

  1. Zhongshan Environment Monitoring Station, Zhongshan 528403, China
  • Received:2009-03-20 Revised:2009-03-20 Online:2009-07-20 Published:2010-10-01

摘要:

研究了连续流动分析仪测定水中阴离子洗涤剂的方法.实验结果表明,在0~2.000mg/L线性区内该方法具有较高的精密度和准确度,其最低检出限为0.010mg/L.与手工亚甲基蓝分光光度法相比,该方法具有自动进样、分析速率快、试剂耗量低等优点,可应用于大批量常规地表水样的分析.

关键词: 连续流动分析仪, 阴离子洗涤剂, 分光光度法

Abstract:

The determination of anionic detergent in water with continuous flowing analyzer has been studied.The experimental results show that the method has higher precision and accuracy within 0-2.000 mg/L anionic detergent in linear area,and the lowest detection limit is 0.010 mg/L.Compared with the manual methylene blue spectrometric method,the advantages of it are auto-sampling,rapid analysis speed and lower reagent consumption,etc.It can be applied to the analysis of large batches of surface water samples.

Key words: continuous flowing analyzer, anionic detergent, spectrometric method

中图分类号: