煤化工废水中无机阴离子的检测方法研究
Study on detection of inorganic anions in coal chemical wastewater
通讯作者:
收稿日期: 2021-03-24
基金资助: |
|
Received: 2021-03-24
作者简介 About authors
高健(1986-),硕士,工程师电话:13612017022,E-mail:
A method of simultaneous separation and detection of F-, Cl-, NO3-, NO2-, Br-, SO42-, PO43- in coal chemical wastewater by ion chromatography(IC) was established. Ion PAC AS14 column was used to separate coal chemical wastewater, with Na2CO3 and NaHCO3 as eluent. The results showed that the detection limits of F-, Cl-, NO3- and SO42- were 0.03, 0.21, 0.55, 0.86 mg/L, respectively. The recovery rates were 95%~104%, and the relative standard deviation(RSD) was less than 3%. The method had the advantages of high sensitivity, simple operation, accurate results and good reproducibility, and was suitable for the detection of inorganic anions in coal chemical wastewater.
Keywords:
本文引用格式
高健, 尚云涛, 张洪杰.
Gao Jian.
1 材料与方法
1.1 仪器设备与试剂
仪器:ICS-600型离子色谱仪,美国戴安;Ion Pac AS14型阴离子分离柱,美国戴安;电导检测器;AERS 500型自动再生抑制器(4 mm),美国戴安;ASDV自动进样器,美国赛默飞世尔;Chromeleon 6型工作站,美国戴安;固相萃取仪,美国色谱科;超纯水机,美国密理博。
试剂:KCl、Na2SO4、NaNO3、NaNO2、NaBr、Na3PO4、NaF、Na2CO3、NaHCO3均为基准试剂,C18固相萃取小柱(500 mg,3 mL),美国色谱科;超纯水电阻率为18.2 MΩ·cm。
1.2 色谱条件
Ion Pac AS14阴离子分离柱(4 mm×250 mm),Ion Pac AG14保护柱(4 mm×50 mm),自动再生抑制器(AERS 500,4 mm),流速为1.2 mL/min,抑制器电流为50 mA,进样体积为10 μL,淋洗液为4.0 mmol/L Na2CO3和1.5 mmol/L NaHCO3。
1.3 样品前处理
样品取自煤制焦工业废水,待测水样经0.45 μm微孔滤膜过滤后备用;分别使用5 mL甲醇、10 mL超纯水活化C18固相萃取柱;将过滤后的样品溶液以2 mL/min的速度注入C18固相萃取柱,弃去前2~ 3 mL样品,收集剩余样品溶液;最后再经0.45 μm微孔滤膜过滤,适当稀释后作为待测样品溶液。
1.4 标准溶液配制
精确配制F-、Cl-、NO2-、Br-、NO3-、PO43-、SO42-混合标准储备液,质量浓度分别为1 000、1 500、1 000、1 000、1 000、1 500、1 500 mg/L,用超纯水定容配制成系列标准溶液。
2 结果与讨论
2.1 色谱体系的建立
Ion Pac AS14型阴离子交换柱适合于多种复杂基质中无机阴离子的快速等度分离。为达到煤化工废水中各种无机阴离子较好的分离效果和分析效率,通过试验探讨了不同淋洗液浓度对色谱体系的影响,结果见表 1。
表1 淋洗液浓度对无机阴离子不对称度和分离度的影响
洗脱液 | 不对称度 | 分离度 | |||||||||||||
F- | Cl- | NO2- | Br- | NO3- | PO43- | SO42- | F- | Cl- | NO2- | Br- | NO3- | PO43- | SO42- | ||
3.0 mmol/L Na2CO3/0.5 mmol/L NaHCO3 | 1.34 | 1.11 | 1.14 | 1.32 | 1.22 | 1.12 | 1.14 | 5.27 | 2.63 | 3.58 | 2.63 | 8.37 | 8.76 | — | |
3.5 mmol/L Na2CO3/1.0 mmol/L NaHCO3 | 1.39 | 1.58 | 1.33 | 1.18 | 1.30 | 1.17 | 1.16 | 4.47 | 2.38 | 3.39 | 2.48 | 6.58 | 6.21 | — | |
4.0 mmol/L Na2CO3/1.5 mmol/L NaHCO3 | 1.35 | 1.30 | 1.39 | 1.11 | 1.35 | 1.18 | 1.13 | 4.31 | 2.27 | 3.45 | 2.20 | 5.56 | 5.67 | — | |
4.5 mmol/L Na2CO3/2.0 mmol/L NaHCO3 | 1.77 | 1.85 | 1.41 | 1.35 | 1.42 | 1.21 | 1.23 | 4.14 | 2.03 | 2.75 | 1.76 | 4.38 | 4.41 | — |
由表 1可知,当淋洗液浓度为3.0 mmol/L Na2CO3/0.5 mmol/L NaHCO3和3.5 mmol/L Na2CO3/1.0 mmol/L NaHCO3时,各无机阴离子分离效果良好,但保留时间较长,分析效率较低;增大淋洗液浓度可以缩短各无机阴离子保留时间,提高分析效率,但当淋洗液浓度过高达到4.5 mmol/L Na2CO3/2.0 mmol/L NaHCO3时会造成色谱峰不对称度增大,峰型受到影响,分离度降低,影响定量分析的准确性。采用4.0 mmol/L Na2CO3/1.5 mmol/L NaHCO3的淋洗液时各无机阴离子不对称度为1.11~1.39,分离度均大于2.0,既达到了较为理想的分离效果又提高了分析效率,适合于煤化工高盐废水中无机阴离子的监测。
7种无机阴离子的色谱图见图 1。
图1
2.2 线性范围及检出限
配制一定浓度梯度的混合标准溶液,按1.2章色谱体系进行分析,以各无机阴离子标准溶液浓度为横坐标,相应无机阴离子色谱峰峰面积为纵坐标建立线性回归方程,计算各组分最低检出限(S/N=3),线性方程、相关性系数及检出限见表 2。
表2 线性回归方程及检出限
阴离子 | 线性方程 | R2 | 线性范围/(mg·L-1) | 检出限/(mg·L-1) |
F- | y=3.249x-0.059 | 0.999 9 | 1~100 | 0.03 |
Cl- | y=4.171x+1.963 | 0.999 6 | 1~150 | 0.21 |
NO2- | y=7.321x+1.201 | 0.999 9 | 1~100 | 0.33 |
Br- | y=8.990x+0.506 | 0.999 5 | 1~100 | 0.52 |
NO3- | y=11.150x+0.519 | 0.999 6 | 1~100 | 0.55 |
PO43- | y=15.672x+1.712 | 0.999 7 | 1~150 | 0.71 |
SO42- | y=6.070x+1.339 | 0.999 5 | 1~150 | 0.86 |
由表 2可知,各无机阴离子组分线性关系良好,R2 > 0.999。
2.3 精密度、重复性及回收率
取某煤化工企业煤制焦工艺废水样品,经处理后进行精密度实验(n=5),计算各无机阴离子保留时间及峰面积相对标准偏差,结果见表 3。
表3 精密度实验
阴离子 | 保留时间RSD/% | 峰面积RSD/% | 阴离子 | 保留时间RSD/% | 峰面积RSD/% | |
F- | 0.17 | 1.25 | NO3- | 0.18 | 1.39 | |
Cl- | 0.15 | 1.46 | PO43- | 0.16 | 1.57 | |
NO2- | 0.13 | 1.32 | SO42- | 1.14 | 1.28 | |
Br- | 0.15 | 1.67 |
取某煤化工企业煤制焦工艺废水样品5份,经处理后进行重复性实验,计算各无机阴离子保留时间及峰面积相对标准偏差,结果见表 4。
表4 重复性实验
阴离子 | 保留时间RSD/% | 峰面积RSD/% | 阴离子 | 保留时间RSD/% | 峰面积RSD/% | |
F- | 1.28 | 1.89 | NO3- | 1.33 | 2.08 | |
Cl- | 1.36 | 1.96 | PO43- | 1.74 | 2.52 | |
NO2- | 1.54 | 2.27 | SO42- | 1.61 | 1.72 | |
Br- | 1.26 | 2.31 |
图2
取某煤化工企业煤制焦工艺废水样品,经处理后进行回收率实验(n=3),先检测样品本底浓度,再加入一定浓度的阴离子标样后检测、计算回收率,结果见表 5。
表5 加标回收率实验
阴离子 | 本底质量浓度/(mg·L-1) | 加标质量浓度/(mg·L-1) | 检测质量浓度/(mg·L-1) | 加标回收率/% |
F- | 3.75 | 2.00 | 5.85 | 102.67 |
3.88 | 2.00 | 5.73 | 96.13 | |
3.61 | 2.00 | 5.71 | 102.77 | |
Cl- | 334.59 | 100.00 | 437.68 | 100.92 |
337.82 | 100.00 | 442.74 | 101.46 | |
336.21 | 100.00 | 435.11 | 99.67 | |
NO2- | 63.18 | 10.00 | 74.29 | 101.76 |
60.27 | 10.00 | 69.24 | 98.29 | |
64.81 | 10.00 | 72.88 | 97.02 | |
Br- | 22.86 | 10.00 | 31.84 | 95.54 |
24.72 | 10.00 | 35.73 | 104.09 | |
26.83 | 10.00 | 35.96 | 96.76 | |
NO3- | 110.43 | 100.00 | 212.62 | 101.98 |
113.65 | 100.00 | 212.35 | 98.86 | |
115.38 | 100.00 | 214.64 | 99.36 | |
PO43- | 394.87 | 100.00 | 495.21 | 100.09 |
396.14 | 100.00 | 495.28 | 99.78 | |
397.22 | 100.00 | 499.41 | 100.55 | |
SO42- | 366.73 | 100.00 | 465.53 | 99.67 |
368.87 | 100.00 | 472.19 | 100.90 | |
365.26 | 100.00 | 466.33 | 100.29 |
图3
3 结论
本研究建立了一种利用离子色谱技术同时分析测定煤化工废水中的7种无机阴离子的方法。结果表明,该方法灵敏度较高,其中F-、Cl-、NO3-、SO42-检出限分别为0.03、0.21、0.55、0.86 mg/L,7种无机阴离子加标回收率为95%~104%,精密度及重现性RSD为0.13%~2.52%。该方法具有分析效率高、线性范围广、重现性好、测定结果准确等优点,可同时进行煤化工废水中7种无机阴离子定量分析,对煤化工废水监测具有重要意义。
参考文献
我国现代煤化工发展的影响因素分析
[J].DOI:10.3969/j.issn.2095-0977.2012.03.011 [本文引用: 1]
/
〈 |
|
〉 |