摘要:
按GB/T16489—1996规定,用分光光度法测定硫化物时,应以氢氧化钠溶液调节pH为10~12的去离子除氧水作为该标准溶液的稀释用水,在加入乙酸锌-乙酸钠溶液后,该溶液在室温下可稳定半年。作者经多次实验认为,用普通去离子水代替上述稀释用水配制的标准溶液,该溶液加入同量的乙酸锌-乙酸钠溶液贮于棕色试剂瓶,在冰箱中密塞保存可稳定半年以上。该法省时省力又不影响分析的准确度,对于广大分析者具有实际意义。
关键词:
硫化物,
稳定性,
分光光度法
Abstract:
According to the regulations of GB / T16489-1996, deionized water containing no oxyen,pH value of which is conditioned to 10 ~ 12 with sodium hydroxide solution, is used for the preparation of standard solution in the determination of sulfide by spectrophotometry. After adding zinc acetate - sodium acetate solution into the above-mentioned standard solution,the lifetime of it is up to half a year at ambient temperature. Deionized water in common use is used in the preparation of standard solution. The experiment results show that the standard solution could be effectively stored within a brown bottle in refrigerator for more than six months by adding the same amount of zinc acetate - sodium acetate solution.The method is of great significance to all the analysts owing to less time and work needed and no influence on its accuracy.
Key words:
sulfide,
stability,
spectrophotometry
中图分类号:
姜晓波, 张素荣, 王丽. 硫化物标准使用液的稳定性实验[J]. 工业水处理, 2003, 23(12): 51-52.
Jiang Xiaobo, Zhang Surong, Wang Li. Stability test of sulfide standard solution[J]. INDUSTRIAL WATER TREATMENT, 2003, 23(12): 51-52.